Using a coil suspension element design for increased shock and vibration resistance, this RTD temperature probe with a 90° bend conforms to the standard 385 curves and comes with spade lugs. It is calibrated and includes an NVLAP-accredited report of calibration from the manufacturer, lab code 200706-0.
Using a coil suspension element design for increased shock and vibration resistance, this RTD temperature probe with a 90° bend conforms to the standard 385 curves and comes with spade lugs. It is calibrated and includes an NVLAP-accredited report of calibration from the manufacturer, lab code 200706-0.
Discontinued!
This product has been discontinued and is no longer available. There are no replacement options.
We can help you find what you need:
Features
Resistance | Nominal 100 Ω |
Temperature Coefficient | 0.004 Ω/Ω/°C Nominal |
Temperature Range | -328° to 788°F (−200 to 420°C) |
Drift Rate (k=2) | ±32°F (0.04°C) at 32° (0°C) after 100 hours at 788°F (420°C) |
Sheath Material | 316 stainless steel |
Leads | PTFE-insulated, nickel-plated stranded copper, 22 AWG |
Termination | Specify. See ordering information |
Time Constant | Four seconds maximum for 63.2% response to step change in water moving at 3 fps |
Immersion | At least 4" (100 mm) recommended |
Calibrated Accuracy (k=2) | ±0.026 at -320°F (-196°C), ±0.046 at 32°F (0°C), ±0.077 at 392°F (200°C), ±0.124 at 788°F (420°C) |
Click on a category to view a selection of compatible accessories with the Fluke 5627A-BEND-S RTD Temperature Probe with 90° bend, spade lugs.
Resistance | Nominal 100 Ω |
Temperature Coefficient | 0.004 Ω/Ω/°C Nominal |
Temperature Range | -328° to 788°F (−200 to 420°C) |
Drift Rate (k=2) | ±32°F (0.04°C) at 32° (0°C) after 100 hours at 788°F (420°C) |
Sheath Material | 316 stainless steel |
Leads | PTFE-insulated, nickel-plated stranded copper, 22 AWG |
Termination | Specify. See ordering information |
Time Constant | Four seconds maximum for 63.2% response to step change in water moving at 3 fps |
Immersion | At least 4" (100 mm) recommended |
Calibrated Accuracy (k=2) | ±0.026 at -320°F (-196°C), ±0.046 at 32°F (0°C), ±0.077 at 392°F (200°C), ±0.124 at 788°F (420°C) |