Featuring a coil suspension element, this precision thermometer/RTD temperature probe delivers a design that尽 shock and vibration resistant. You can use your DIN/IEC RTD meter fully, since the unit conforms to standard 385 curves. With a mineral-insulated sheath that measures 6in x 0.19in and banana plug connections, this thermometer provides a highly flexible design.
Featuring a coil suspension element, this precision thermometer/RTD temperature probe delivers a design that尽 shock and vibration resistant. You can use your DIN/IEC RTD meter fully, since the unit conforms to standard 385 curves. With a mineral-insulated sheath that measures 6in x 0.19in and banana plug connections, this thermometer provides a highly flexible design.
It is an excellent value for a precision thermometer and it has the price-to-accuracy and price-to-durability ratios you should demand in all of the precision thermometers you buy.
Features
Resistance | Nominal 100Ω |
Temperature Coefficient | 0.00385Ω/Ω/°C nominal |
Temperature Range | -328 to 788°F (-200 to 420°C) (5627A-6 and 5627A-9 to 300°C; transition and cable temperature: 0 to 150°C) |
Drift Rate (k=2) | ±0.04°C at 0°C after 100 hours at 420°C |
Sheath Material | 316 stainless steel |
Leads | PTFE-insulated, nickel-plated stranded copper, 22AWG |
Termination | See ordering information |
Time Constant | 4 seconds maximum for 63.2% response to step change in water moving at 3fps |
Bending Radius | Sheath may be ordered with a bend on a minimum radius of 0.75in (19mm) except for 2in (50mm) area of sheath near tip; Fluke lab requires 8in (20cm) of unbent sheath to re-calibrate |
Calibration | Includes manufacturer's NVLAP-accredited (lab code 200706-0) calibration and table with R vs T values in 1°C increments from -196 to 500°C (to 300°C for 5627A-6 and 5627A-9) ITS-90 coefficients included |
Immersion | At least 4in (100mm) recommended |
Calibrated Accuracy (Includes Calibration Uncertainty and 100-Hour Drift) (k=2) | ±0.026°C at -196°C ±0.046°C at 0°C ±0.077°C at 200°C ±0.124°C at 420°C |
Size | 6 x 0.19in (152 x 4.7mm) |
Click on a category to view a selection of compatible accessories with the Fluke 5627A-6-J Precision Thermometer/RTD Temperature Probe with Banana Plugs, 6 x 0.19in, -200 to 300°C.
Resistance | Nominal 100Ω |
Temperature Coefficient | 0.00385Ω/Ω/°C nominal |
Temperature Range | -328 to 788°F (-200 to 420°C) (5627A-6 and 5627A-9 to 300°C; transition and cable temperature: 0 to 150°C) |
Drift Rate (k=2) | ±0.04°C at 0°C after 100 hours at 420°C |
Sheath Material | 316 stainless steel |
Leads | PTFE-insulated, nickel-plated stranded copper, 22AWG |
Termination | See ordering information |
Time Constant | 4 seconds maximum for 63.2% response to step change in water moving at 3fps |
Bending Radius | Sheath may be ordered with a bend on a minimum radius of 0.75in (19mm) except for 2in (50mm) area of sheath near tip; Fluke lab requires 8in (20cm) of unbent sheath to re-calibrate |
Calibration | Includes manufacturer's NVLAP-accredited (lab code 200706-0) calibration and table with R vs T values in 1°C increments from -196 to 500°C (to 300°C for 5627A-6 and 5627A-9) ITS-90 coefficients included |
Immersion | At least 4in (100mm) recommended |
Calibrated Accuracy (Includes Calibration Uncertainty and 100-Hour Drift) (k=2) | ±0.026°C at -196°C ±0.046°C at 0°C ±0.077°C at 200°C ±0.124°C at 420°C |
Size | 6 x 0.19in (152 x 4.7mm) |
Click on a category to view a selection of compatible accessories with the Fluke 5627A-6-J Precision Thermometer/RTD Temperature Probe with Banana Plugs, 6 x 0.19in, -200 to 300°C.