Capable of withstanding temperatures from -100 to 156°C, this precision RTD freezer probe is ideal for verifying the temperature of freezers or autoclaves when a thermo-well isn腎 available. The probe prevents moisture infiltration thanks to its sealed construction, and it can be completely immersed in liquids. The unit features gold pin connections.
Capable of withstanding temperatures from -100 to 156°C, this precision RTD freezer probe is ideal for verifying the temperature of freezers or autoclaves when a thermo-well isn腎 available. The probe prevents moisture infiltration thanks to its sealed construction, and it can be completely immersed in liquids. The unit features gold pin connections.
Available in a 0.25in (6.35mm) diameter x 6in (125mm) long Inconel sheath. With accuracy, which includes calibration uncertainty and short-term drift, of ±0.05°C over its full range, the 5623B is just right as a secondary standard for calibration of other process sensors.
Features
Resistance | Nominal 100Ω (±0.1Ω) |
Temperature Coefficient | 0.003925Ω/Ω/°C nominal |
Temperature Range | -148 to 312.8°F (-100 to 156°C) |
Transition Temperature | -148 to 312.8°F (-100 to 156°C) |
Drift Rate | ±0.01°C per year maximum at 0°C, when used periodically at max temperature |
Sheath Material | Inconel 600 |
Leads | PTFE-insulated, silver-plated stranded copper, 22AWG |
Termination | Specify (see models and accessories) |
Calibration | Includes manufacturer's NVLAP-accredited, lab code 200706-0, calibration and table with R vs. T values in 1°C increments from -80°C to 156°C; ITS-90 coefficients included |
Calibrated Accuracy (Includes Calibration Uncertainty And Drift) (k=2) | ±0.05°C over the full range |
Cable Length | 20ft (6.1m) |
Size | 0.25 x 6in (6.35 x 152mm) |
Click on a category to view a selection of compatible accessories with the Fluke 5623B-6-G Precision RTD Freezer Probe with Gold Pins, 6 x 0.25in, -100 to 156°C.
Resistance | Nominal 100Ω (±0.1Ω) |
Temperature Coefficient | 0.003925Ω/Ω/°C nominal |
Temperature Range | -148 to 312.8°F (-100 to 156°C) |
Transition Temperature | -148 to 312.8°F (-100 to 156°C) |
Drift Rate | ±0.01°C per year maximum at 0°C, when used periodically at max temperature |
Sheath Material | Inconel 600 |
Leads | PTFE-insulated, silver-plated stranded copper, 22AWG |
Termination | Specify (see models and accessories) |
Calibration | Includes manufacturer's NVLAP-accredited, lab code 200706-0, calibration and table with R vs. T values in 1°C increments from -80°C to 156°C; ITS-90 coefficients included |
Calibrated Accuracy (Includes Calibration Uncertainty And Drift) (k=2) | ±0.05°C over the full range |
Cable Length | 20ft (6.1m) |
Size | 0.25 x 6in (6.35 x 152mm) |