With a wide temperature range and high-purity platinum element, this secondary reference temperature standard is ideal for lab or field calibration. The temperature standard measures 12 x 0.25in and has bare wire connections.
With a wide temperature range and high-purity platinum element, this secondary reference temperature standard is ideal for lab or field calibration. The temperature standard measures 12 x 0.25in and has bare wire connections.
Discontinued!
This product has been discontinued and is no longer available.
This is the direct replacement:
It bridges the gap between a 100Ω industrial RTD and an SPRT.
Features
Temperature Range | -328 to 572°F (-200 to 300°C) |
Nominal Resistance At 0°C | 100Ω ±0.10Ω |
Temperature Coefficient | 0.0039250Ω/Ω/°C |
Accuracy (Includes Calibration And 100-Hour Drift (k=2)) | ±0.024°C at -200°C; ±0.012°C at 0°C; ±0.035°C at 420°C |
Drift (After 100 Hours At Max Temp, 95% Confidence (k=2)) | ±0.007°C at 0.010°C |
Sensor Length | 1.1in (28mm) |
Sensor Location | 0.27in ±0.13in (6.9mm ±3.3mm) from tip |
Lead-Wire Cable Type | PTFE insulated with PTFE jacket, 22AWG |
Lead-Wire Length | 6ft (183cm) |
Calibration | NVLAP-accredited calibration included, lab code 200706-0 |
Click on a category to view a selection of compatible accessories with the Fluke 5616-12-B Secondary Reference Temperature Standards with Bare Wire, 12 x 0.25in.
Temperature Range | -328 to 572°F (-200 to 300°C) |
Nominal Resistance At 0°C | 100Ω ±0.10Ω |
Temperature Coefficient | 0.0039250Ω/Ω/°C |
Accuracy (Includes Calibration And 100-Hour Drift (k=2)) | ±0.024°C at -200°C; ±0.012°C at 0°C; ±0.035°C at 420°C |
Drift (After 100 Hours At Max Temp, 95% Confidence (k=2)) | ±0.007°C at 0.010°C |
Sensor Length | 1.1in (28mm) |
Sensor Location | 0.27in ±0.13in (6.9mm ±3.3mm) from tip |
Lead-Wire Cable Type | PTFE insulated with PTFE jacket, 22AWG |
Lead-Wire Length | 6ft (183cm) |
Calibration | NVLAP-accredited calibration included, lab code 200706-0 |
Click on a category to view a selection of compatible accessories with the Fluke 5616-12-B Secondary Reference Temperature Standards with Bare Wire, 12 x 0.25in.